Rotors for CC40NX series

CC40NX

This model is designed expressly to purify large-volume samples such as viral particles or nanosized particles. Its high performance is defined by the 8.0L maximum capacity with batch centrifugation and 7.7L maximum capacity with continuous flow centrifugation, speeds up to 40,000rpm and the maximum RCF of 118,000xg with Titanium alloy made core.

CC40SNX

With approximately one-half the capacity of CC40NX, this model is ideal for research, development, pilot production and small-volume production of pharmaceuticals and nanosized materials. It has a maximum rotor capacity of 1.6L. max. speed of 40,000rpm and maximum RCF of 118,000xg with Titanium alloy made core.

Specifications of Rotors and Cores

for CC40NX

Rotor model C40CT4*1 C40CT3
Type of core Core(H)
(high-flow type)
Ti Core(H)
(high-flow type)
Core(D) Ti Core(D) Core(E)*2 Core(F) Core(G)
External appearance
of core
*3 *4
Max. speed 40,000rpm 40,000rpm 40,000rpm 40,000rpm 36,000rpm 40,000rpm 40,000rpm
Max. RCF 118,000xg 118,000xg 118,000xg 118,000xg 96,000xg 118,000xg 118,000xg
Rmax 66mm 66mm 66mm 66mm 66mm 66mm 66mm
Rotor weight 31kg 31kg 31kg 31kg 31kg 31kg 31kg
Rotor material Titanium
alloy
Titanium
alloy
Titanium
alloy
Titanium
alloy
Titanium
alloy
Titanium
alloy
Titanium
alloy
RCF at Rmin 97,500xg 97,500xg 97,500xg 97,500xg 37,600xg 97,500xg 47,600xg
Rmin 54.5mm 54.5mm 54.5mm 54.5mm 21mm 54.5mm 26.6mm
Core length 76.1cm 76.1cm 76.1cm 76.1cm 76.1cm 76.1cm 76.1cm
Core material Noryl™
Regin
Titanium
alloy
Noryl™
Regin
Titanium
alloy
Noryl™
Regin
Noryl™
Regin
Noryl™
Regin
Capacity 3.2L 3.2L 3.2L 3.2L 8.0L 3.2L
(preclarifier: 0.3L)
7.7L
O-ring material FKM*5 FKM*5 FKM*5 FKM*5 FKM*5 FKM*5 FKM*5
Applications Density gradient
centrifuga-
tion of virus,
liposome and other
nanosized particles etc.
(for high viscous sample)
Density gradient
centrifuga-
tion of virus,
liposome and other
nanosized particles etc.
(for high viscous sample)
Density gradient
centrifuga-
tion of virus,
liposome and other
nanosized particles etc.
Density gradient
centrifuga-
tion of virus,
liposome and other
nanosized particles etc.
Batch centrifuga-
tion
of HBsAg or nanosized
particles etc.
Density gradient centrifuga-
tion
of virus, liposome and other
nanosized particles etc.
The preclarifier core permits
separation of cells and
aggregation substance.
Rough separation of
large-volume samples
for pelleting
  1. The C40CT4 rotor assemble, using Core (H) or Ti Core (H), is designed to reduce a pressure in sample feed line at continuous centrifugation by about 30% lower than the one with the C40CT3 rotor assemble using Core (D) or Ti Core (D). It is suitable for purification of large-volume or viscous samples.
  2. The Core (E) is not for continuous flow centrifugation but for batch centrifugation.
  3. The shape and size are the same as Core (H)
  4. The shape and size are the same as Core (D)
  5. FKM complies to USP Class IV.
  • The name of actual companies and products mentioned herein may be the trademarks of their respective owners.

for CC40SNX

Rotor model C40CTS4*1 C40CTS3
Type of core Core (AH)
(high-flow type)
Ti Core (AH)
(high-flow type)
Core (A) Ti Core (A) Core (B)
External appearance
of core
*2 *3
Max. speed 40,000rpm 40,000rpm 40,000rpm 40,000rpm 40,000rpm
Max. RCF 118,000xg 118,000xg 118,000xg 118,000xg 118,000xg
Rmax 66mm 66mm 66mm 66mm 66mm
Rotor weight 15kg 15kg 15kg 15kg 15kg
Rotor material Titanium alloy Titanium alloy Titanium alloy Titanium alloy Titanium alloy
RCF at Rmin 97,500xg 97,500xg 97,500xg 97,500xg 97,500xg
Rmin 54.5mm 54.5mm 54.5mm 54.5mm 54.5mm
Core length 382mm 382mm 382mm 382mm 382mm
Core material Noryl™ Regin Titanium alloy Noryl™ Regin Titanium alloy Noryl™ Regin
Capacity 1.6L 1.6L 1.6L 1.6L 1.6L
(preclarifier: 0.15L)
O-ring material FKM*4 FKM*4 FKM*4 FKM*4 FKM*4
Applications Density gradient centrifugation
of virus, liposome and other
nanosized particles etc.
(for high viscous sample)
Density gradient centrifugation
of virus, liposome and other
nanosized particles etc.
(for high viscous sample)
Density gradient centrifugation
of virus, liposome and other
nanosized particles etc.
(for high viscous sample)
Density gradient centrifugation
of virus, liposome and other
nanosized particles etc.
(for high viscous sample)
Density gradient centrifugation
of virus, liposome and other
nanosized particles etc.
The preclarifier core permits separation
of cells and aggregation substance.
  1. The C40CTS4 rotor assemble, using Core (AH) or Ti Core (AH), is designed to reduce a pressure in sample feed line at continuous centrifugation by about 30% lower than the one with the C40CTS3 rotor assembly using Core (A) or Ti Core (A). It is suitable for purification of large-volume or viscous samples.
  2. The photo is Core (AH). The shape and size are the same as Core (AH)
  3. The photo is Core (A). The shape and size are the same as Core (A)
  4. FKM complies to USP Class IV.
  • The name of actual companies and products mentioned herein may be the trademarks of their respective owners.

Please feel free to contact our sales representatives to offer appropriate combination of the rotor and the core based on your application.